Immunophenotyping: Instrument Calibration and Reagent Qualification for Immunophenotyping Analysis of Human Peripheral Blood Mononuclear Cell Cultures

Immunophenotyping: Instrument Calibration and Reagent Qualification for Immunophenotyping Analysis of Human Peripheral Blood Mononuclear Cell Cultures

Standard Number ITA-37.1
Organization Nanotechnology Characterization Laboratory USA
Level National
Category Test Method | Characterization | Measurement
Status
  • OCT 2022 Published
ABSTRACT

This panel includes antibody-fluorophore conjugates that allow for analysis of different lymphocyte populations including B cells and T cells (CD8+ T cells, CD4+ T cells, regulatory T (Treg) cells, naïve T cells, and γδ TCR T cells). This panel also determines cellular CD25 and CD154 expression which are markers of proliferation and co-stimulation/presentation, respectively.

When used in conjunction with other immunoassays, this protocol aids in establishing efficacy and safety profiles of engineered nanoparticles used for vaccine or drug delivery. This protocol has two parts, ITA-37.1 described in this document and intended for instrument calibration, and ITA-37.2, described in a separate document and intended for the analysis of nanoparticle-treated cells.